Antegrade screw fixation of a scaphoid fracture

1. General considerations

Introduction

For the fixation of waist or proximal pole fractures of the scaphoid, headless compression screws (2.4 or 3.0 mm) may be used in an antegrade fashion.

Scaphoid fracture – Headless screw fixation
Note on illustrations

In this basic technique, associated bony and ligamentous injuries are not shown for clarity.

Choice of implant

If the proximal pole fragment is larger than 5 mm, a cannulated, headless, self-compressing screw (2.4 or 3.0 mm) is the implant of choice.

For smaller proximal pole fragments, the use of a mini headless bone screw is advisable.

K-wires may be a better option for fixation of very small fragments (flakes).

Scaphoid proximal pole fractures - choice of implant

Scapholunate ligament reattachment

If a bony avulsion fracture of the scapholunate ligament from the scaphoid is present, fixing the fragment, either with K-wires, or a small screw (1.5 mm), is necessary.

Scaphoid fracture - Scapholunate ligament reattachment

Comminuted scaphoid fractures

Unlike the treatment of an isolated comminuted scaphoid fracture, an associated scaphoid fracture should be approached through the same dorsal approach as the other injuries.

In such a case, adding a K-wire or a second screw may help improve fracture stabilization.

Comminuted scaphoid fracture fixed with a headless compression screw

Anatomical considerations

80% of the surface of the scaphoid is covered with articular cartilage. This greatly limits potential points of entry for fixation devices.

An additional constraint is the curved shape of the scaphoid.

This means that a wire or fixation device along the true central axis of the scaphoid is not possible from a palmar approach. Occasionally, access to a distal entry point for a device can only be gained by a limited excavation of the edge of the trapezium.

The scaphoid

Fracture displacement forces

In displaced fractures of the waist of the scaphoid, the distal pole tends to rotate into flexion in relation to the proximal pole, the lunate, and the triquetrum, which lie in extension. This can create a rotational and angular deformity at the fracture site – the so-called “humpback deformity.”

Scaphoid waist fracture - fracture displacement forces

Preoperative planning

Conventional radiographs do not adequately demonstrate the complete fracture configuration. A CT scan is recommended to reveal the degree of displacement.

Scaphoid fracture – CT scan

Before starting the surgical procedure, reexamine the fracture pattern under the image intensifier. Ensure the fracture is suitable for a percutaneous technique and that no secondary displacement has occurred. This is especially indicated for displaced or oblique fractures.

Patient in supine position with arm on radiolucent side table to allow use of C-arm.

2. Reduction

Inspection of the fracture

Look for additional lesions, especially scapholunate ligament ruptures.

The picture shows a proximal pole fracture combined with a complete scapholunate rupture. The head of the capitate is visible deeper in the wound. The proximal pole fragment has been delivered into the wound and has no soft-tissue attachments to the remaining carpus. Its vitality is therefore in question.

Scaphoid fracture – inspection of the fracture

Reduction

If the fracture is displaced, reduce it with small pointed reduction forceps.

Scaphoid fracture – reduction

3. Guide-wire insertion

The entry point is at the proximal pole, directly adjacent to the scapholunate ligament insertion.

Insert the guide wire in the axis of the shaft of the first metacarpal, in radial abduction.

Scaphoid waist fracture – guide-wire insertion

Confirm accurate advancement of the guide wire in the scaphoid axis with an image intensifier in at least two planes and perpendicular to the fracture plane.

Do not penetrate the scaphotrapezial joint with the guide wire.

Scaphoid fracture – guide-wire insertion: confirmation of position with image intensifier

4. Screw insertion

Insert the headless compression screw in a standard manner.

Select a screw 2–4 mm shorter than the measured length with the appropriate thread length. In most cases, a 16–20 mm cannulated screw is the appropriate length.

Pitfall: A long screw can impinge upon the scaphotrapezial joint and lead to articular damage of both bones, resulting in osteoarthritis.
Scaphoid fracture – headless compression screw insertion

Confirming screw position

Check the final position of the screw and the scaphoid stability with an image intensifier.

Scaphoid fracture - screw fixation
Go to diagnosis